Defaultsettings
SettingDescription
ED
ICUNICUORPACU
PR(SpO
2
)alarm
limits
Selectthe
high/lowalarm
limits.
150/50
(Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80)
200/90
(Infant
prole:
180/80)
160/40
(Neuro
prole:
120/50;
Adolescent
prole:
150/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
150/50
(Neuro
prole:
120/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
PR(SpO
2
)alarm
on
Enablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
PR(UAC)alarm
limits
Selectthe
high/lowalarm
limits.
150/50200/90
(Infant
prole:
180/80)
150/50
PR(UAC)alarm
on
Enablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
PR(Art)alarm
limits
Selectthe
high/lowalarm
limits.
150/50
(Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80)
200/90
(Infant
prole:
180/80)
160/40
(Neuro
prole:
120/50;
Adolescent
prole:
150/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
150/50
(Neuro
prole:
120/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
PR(Art)alarmonEnablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
PR(Fem)alarm
limits
Selectthe
high/lowalarm
limits.
150/50
(Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80)
200/90
(Infant
prole:
180/80)
160/40
(Neuro
prole:
120/50;
Adolescent
prole:
150/50;
Pediatric
prole:
170/70;
Infant
150/50
(Neuro
prole:
120/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
2094480-001CARESCAPEONE