Defaultsettings
SettingDescription
NICU
PrimaryHRSource
SelecttheparametertocalculateHR.
AUTO
(USonly:ECG)
ShowPulseRateEnablethepulseratedisplay.Disabled
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>UAC1>Advancedto
adjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefault
settings.
SettingDescription
NICU
FilterHzSelectthewaveformlter.
12
CatheterDisconnectEnablethedisconnectalarm.Enabled
PrimaryArterialSourceSelecttheprimaryarterialsource.
AUTO
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>UAC1>Alarms>UAC
1>Alarmstoadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhe
factorydefaultsettings.
SettingDescription
NICU
Systolicalarmlimits
Selectthehigh/lowalarmlimits.90/40
(CardiacandFull-termproles:
100/40;Infantprole:100/50.)
Systolicalarmson/off
Enablealarmwhenlimitsareviolated.Enabled
Meanalarmlimits
Selectthehigh/lowalarmlimits.60/30
(Cardiac,Full-termandInfant
proles:70/40.)
Meanalarmson/off
Enablealarmwhenlimitsareviolated.Enabled
Diastolicalarmlimits
Selectthehigh/lowalarmlimits.50/20
(CardiacandFull-termproles:
60/20;Infantprole:60/30.)
Diastolicalarmson/off
Enablealarmwhenlimitsareviolated.Disabled
(Infantprole:Enabled)
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>UAC1>Alarms>
UAC1>Alarms>PR(UAC1)Alarmstoadjusttheheartratealarmdefaultsettingsforthe
currentpatient.
CVP2pressuresitedefaultsettings
ThedefaultP2pressuresitefortheED,ICU,OR,andPACUsoftwarepackageisCVP.Select
MonitorSetup>MainSetup>ParameterSetup>Page1>CVP2>Setup>Page1or
Page2toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactory
defaultsettings.
362CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017