Defaultsettings
SettingDescription
ED
ICUORPACU
LabelSelectthepressure
sitelabel.
CVP
ScalemmHgSelectthewaveform
scale.
0-20
(Infantprole:0-10)
HemoSweep
Speed
Selectthewaveform
displaysweepspeed.
25mm/s
DisplayFormatSelectthepressure
valuesdisplayformat.
Meanonly
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CVP2>Advancedto
adjusttheFilterHzsettingforthecurrentpatient.Thefactorydefaultsettingis12.
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CVP2>Alarmstoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUORPACU
Meanalarmlimits
Selectthehigh/low
alarmlimits.
15/0
(Infantprole:10/0)
Meanalarms
on/off
Enablealarmwhen
limitsareviolated.
Enabled
UVC2pressuresitedefaultsettings
ThedefaultP2pressuresitefortheNICUsoftwarepackageisUVC.SelectMonitorSetup>
MainSetup>ParameterSetup>Page1>UVC2>Setup>Page1orPage2toadjustthese
settingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
NICU
LabelSelectthepressuresitelabel.
UVC
ScalemmHgSelectthewaveformscale.
0-10
HemoSweepSpeedSelectthewaveformdisplaysweep
speed.
25mm/s
DisplayFormatSelectthepressurevaluesdisplay
format.
Meanonly
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>UVC2>Advancedto
adjusttheFilterHzsettingforthecurrentpatient.Thefactorydefaultsettingis12.
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>UVC2>Alarmstoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
NICU
Meanalarmlimits
Selectthehigh/lowalarmlimits.5/0
(Infantprole:10/0)
Meanalarmson/off
Enablealarmwhenlimitsareviolated.Enabled
2094480-001CARESCAPEONE363