Systemintroduction
WARNING
PHYSICALINJURY.Donotinstallthedeviceaboveapatient.
Makesurethebatteryiscompletelyinsertedandthebattery
dooriscompletelyclosed.Fallingbatteriescouldseriouslyor
fatallyinjureneonatalorothervulnerablepatients.
WARNING
EXPLOSIONHAZARD.Donotincinerateabatteryorstoreat
hightemperatures.Seriousinjuryordeathcouldresult.
Batteriesmustbeinstalledandfullychargedpriortotakingintouse.
Testingthebatterycharge
Beforeinstallingabattery,verifythebattery’sstateofcharge.Eachbatterymustbe
fullychargedbeforeuse.
1.PresstheTESTbuttononthebatteryandcheckthegreencharginglevelindicators
toseehowmuchchargeisleft:
●FourLEDsilluminated:75%to100%offull-chargecapacity.
●ThreeLEDsilluminated:50%to74.9%offull-chargecapacity.
●TwoLEDsilluminated:25%to49.9%offull-chargecapacity.
●OneLEDilluminated:11%to24.9%offull-chargecapacity.
●OneLEDashing:<11%offull-chargecapacity.
InsertingandremovingtheCARESCAPEONEbattery
WARNING
EXPLOSIONORFIRE.Usingnon-recommendedbatteries
couldresultininjury/burnstothepatientsorusers.Onlyuse
batteriesrecommendedormanufacturedbyGE.Thewarranty
canbevoidedifnon-recommendedbatteriesareused.
WARNING
PHYSICALINJURY.Donotinstallthedeviceaboveapatient.
Makesurethebatteryiscompletelyinsertedandthebattery
dooriscompletelyclosed.Fallingbatteriescouldseriouslyor
fatallyinjureneonatalorothervulnerablepatients.
WARNING
PHYSICALINJURY.Donotinstallthedeviceaboveapatient.
Leaksfromthebatterycellscanoccurunderextreme
conditions.Theliquidiscaustictotheeyesandskin.Ifthe
liquidcomesincontactwitheyesorskin,ushwithclean
waterandseekmedicalattention.
CAUTION
LOSSOFMONITORING.Topreventlossofmonitoring,only
changetheCARESCAPEONEbatterywhenCARESCAPEONEis
connectedtoapoweredCARESCAPEDockF0.
2094480-001CARESCAPEONE
43
CARESCAPE ONE User Manual DRAFT 26 October 2017