Messages
Message
Location
ExplanationWhattodo
●Error1:PMCupdate
failed.Turnthe
monitoroffandthen
onagain.
●start-up
screen
UnsuccessfulPMCupdate.●Turnthemonitoroffandthen
onagain.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Error2:PMCupdate
requiresamains
supply.Pluginthe
powercordandturn
themonitoroffand
thenonagain.
●start-up
screen
UnsuccessfulPMCupdate.●Pluginthepowercordand
turnthemonitoroffandthen
onagain.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Externalalarmlight
disconnect.Check
USBconnection.
●al.areaThereisacommunicationerror.●Waitandseeifthemessage
disappears.
●Tryselectingaudiopause
andseeifthemessage
disappears.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●IdenticalIPaddress
noticed
●al.areaTwoormoremonitorshavethe
sameIPaddress.
●Contactauthorizedservice
personnel.
●Identicalunit&bed
namenoticed
●al.areaTwoormoremonitorshavethe
sameunitandbedname.
●Contactauthorizedservice
personnel.
●IncompatibleModule●al.areaThemoduleisnotcompatible.●Replacewithcompatible
module.SeetheCARESCAPE
ModularMonitors
SupplementalInformation
Manual.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Incompatible
TramRAC
●al.areaIncompatibleTram-Rac
connected.
●Replacewithcompatible
Tram-Racmodule.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Incorrectbarcode
value
●al.areaThebarcodestringdiffersfrom
thedenedvalues.
●Contactauthorizedservice
personnel.
●InformationCenter
busy
●al.area
S/5Networkhastoomany
bed-to-bedconnections.
●Tryagainlater.
●InformationCenter
notresponding
●al.area
S/5Networkisbusyandthe
centralstationisnotresponding.
●Tryagainlater.
●Invalidbarcode
conguration
●al.areaThebarcodecongurationisnot
correct.
●Contactauthorizedservice
personnel.
508CARESCAPEModularMonitors2062971-001