EasyManua.ls Logo

GE CT-9800 - Page 133

GE CT-9800
510 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-10-11
R
INSTALL ‘CR’
* * * LOADING INSTALLATION WORK FILES
STATION TYPE (1=SCAN 2=INDEP): (The station you are presently at)
DOES YOUR SYSTEM HAVE AN HTD OPTION (YIN)
XENON OR HILIGHT DETECTOR (X/H):
HILIGHT DETECTOR ID NUMBER: XXXX (4 DIGITS)
HILIGHT DETECTOR ASSEMBLY NUMBER: (12 DIGITS, NO -, 46 NUMBER)
HOSPITAL NAME: (24 CHARACTERS MAX)
UNIQUE SYSTEM NUMBER: (SYSTEM ID NUMBER)
PREFERRED PHYSICIAN TITLE, DIAGNOSTIC IAN OR RADIOLOGIST (D/R):
LINE PRINTER? (Y/N):
OPTICAL DISK PRESENT (Y/N)
TAPE DENSITY (800/1600): [1600] (archive tape density)
MAG TAPE LENGTH (600/1600/2400): [2400] (large reel tape size)
IMAGE COMPRESSION TO TAPE (Y/N): [Y] (Image compression stores more images per
tape)
STARTING ARCHIVE TAPE ID: (1-32699. Check patient log book for last
number.)
DISK TYPE (1 = ARGUS, 2 = ACE : 1) (If the system is configured with a Zebra disk
drive or HTD option, this question will not appear.
If the system has an Argus. enter 1; If the
system has an Ace, enter 2)
2 DISK DRIVE SYSTEM (Y/N): (If the system is configured with ACE disk
drive(s) this question will not appear)
BONE MINERAL DENSITOMITRY AT THIS SITE?(Y/N):
DATALINK PRESENT (Y/N): (If N, no further DataLink questions asked)
NUMBER OF STATIONS: n (1-26) (Each DataLink box contains a dip switch
which identifies station or node number.
Numbering is sequential from 1 to 26. “O” is not
used and no numbers are skipped. Backup packs
must have same Station ID).
STATION ID (1-n): (ID’1 of station you are loading software on.)
IMAGE TAPE STATION ID (1-n):(The ID of station where image storage is usually
performed. Refer to DataLink Section of the
Operator Manual for more details, if desired).