GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 24 Direction 18000
4-10-12
STATION 1 (This sequence of questions is repeated n times)
STATION TYPE (1=SCAN 2=INDEP): (Is Station 1 a Scanner or IC?)
. .
. .
STATION n
STATION TYPE (1=SCAN 2=INDEP):
(Is Station n a Scanner or IC?)
DIAGNOSTIC CONSOLE ON YOUR SYSTEM (Y/N):(Only asked if scan station type=1 selected)
X-RAY TUBE TYPE (Metal Tubes not allowed)
1)HIGH PERFORMANCE WITH CUPD3212T or D3202T
2) HIGH PERFORMANCE WITHOUT CUP D3222T
3)STANDARDD1192JR
X-RAY TUBE TYPE (1/2/3): (Enter appropriate selection)
NOTE: IF YOU HAVE D3202T SELECT 1) HIGH PERFORMANCE WITH CUP.
WHEN INSTALL IS COMPLETE, LOAD FILE 6 FROM THE PROPRIETARY DIAGNOSTIC TAPE,
AND THEN RUN REGEN TO SELECT THE PROPER TUBE.
NUMBER OF SCAN DATA FILES: For Scanner For IC with AP
24-57 1-57 (1 ZEBRA drive)
24-100 1-100 (2 ZEBRA drives or
(1-2 ARGUS drives
or
(1-2 ACE drives)
For IC with no AP, enter 0
STARTING EXAM NUMBER: (1-32000) (Only asked if scan station type selected; sets the
minimum value for REGEN)
SCAN DATA FILE SIZE (S=STANDARD L=LARGE/OVERSCAN): Since overscan files are very large,
check with the customer and see if
they will be performing overscans.
Otherwise, select Standard.
DOES YOUR SYSTEM HAVE A QUICK 3D OPTION (Y/N):
***ALLOCATING SWAP AREA FOR THE SBC OPERATING SYSTEM
(If HTD Option)
***RESTORING THE SBC OPERATING SYSTEM (If HTD Option)
(Much tape and disk activity)
PLEASE REMOVE TAPE #1 AND LOAD TAPE #2 (Do this while system is generating SLCF files)
BAD BLOCKS FOR DAO (DZ0):
THERE ARE NO BAD BLOCKS ON THE DISK.