Service Menu
47
Document no. M1137266-02
HW error Module hardware error status. See Table 7 Module hardware error status for a detailed
description of the message.
Table 7 Module hardware error status
Testbit Testbit is a module production phase test status and is always 0000 (HEX). If the testbit status is
0000, then the module should be returned to the factory.
Voltages Vdd; 5 V digital units driving voltage received from the module frame.
5vdc; 5 Volts of the module derived internally from 15vdc.
15vdc; 15 Volts of the module received from the module frame.
Pneuma voltage; 15 V voltage driving pneumatic elements of the module, derived from 15V
dirty.
15V dirty; 15 V voltage received from the module frame.
bit 0 Lamp error
bit 1, 2, 3 not used
bit 4 CPU EEPROM error
bit 5 Sensor EEPROM error
bit 6 Internal AD converter error
bit 7 External AD converter error
bit 8 CO
2
termopile error
bit 9 Reserved
bit 10 Temperature detector error
bit 11 Pump pressure sensor error
bit 12 Catheter pressure sensor error
bit 13 Pneumatic power failure
bit 14-15 Not used