Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
70
Document no. M1137266-02
4.16.5 DELP service menu
DELP IBP
Systolic Pressure Variation (SPV) and delta Pulse
Pressure (dPP) IBP data has been collected in the
DELP service page.
Measurement: Delta pressure measurement is
always Present (Yes) and Active (Yes) in Monitor
software 07.
Invasive BP:
Channel is always 0 in Monitor software 07.
Beats is the number of Art pulses detected by the
calculation algorithm since monitor startup.
Arrhytmh is the number of detected artefact
cycles.
Timeouts is the number of missing heart beats.
SBP is the systolic pressure beat-by-beat value
(mmHg x 100).
DBP is the diastolic pressure beat-by-beat value (mmHg x 100).
SPV:
Breaths is the number of breaths detected from the Art waveform variation for the SPV calculation.
Arrhytmh is the number of detected artefact cycles in SPV calculation.
Inv is the count of invalid (e.g. resp rate too high) breath cycles.
Beats / Br is the average of Art pulses detected during one breath for the SPV calculation.
1 min ct is the detected respiration rate for SPV calculation.
Cycle count is the count of breaths used in SPV calculation.
Breath d. is the breath mean duration (seconds x 100).
dPP:
Breaths is the number of breaths detected from the Art waveform variation for the dPP calculation.
Arrhytmh is the number of detected artefact cycles in dPP calculation.
Inv is the count of invalid (e.g. resp rate too high) breath cycles.
Beats / Bris the average of Art pulses detected during one breath for the dPP calculation.
1 min ct is the detected respiration rate for SPV calculation.
Cycle count is the count of breaths used in dPP calculation.
Breath d. is the breath mean duration (seconds x 100).