Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
14
Document no. M1125635-05
8. Service reset -switch
Check that the service reset -switch is intact. Press the switch at least for five seconds.
"
9. Equipotential tap and rear panel screws
Check that the equipotential tap and all the rear panel screws are tightened properly.
"
PC boards
10. PC board screws
Check that all the rear side PC boards are secured to the Central Unit with screws.
"
11. PC board connectors
Check that the connectors on the rear panels of the PC boards are clean and intact.
"
12. Block screws for cables
Check that all block screws for cables are in place and tightened properly. Check also that
their threads are intact.
"
13. Grounding plates
Check that the grounding plates under the PC board rear panels are attached properly
and are not bent.
"
14. Optional display controller board B-DISPX jumpers/DIP switches
If available, disconnect the display controller board(s) and check that the address and
interruption jumpers/ DIP switches have been set correctly. The instructions for setting
the jumpers can be found in the “Displays and Display Controllers” slot of the Technical
Reference Manual,
Leave the board(s) disconnected.
"
Disconnect the UPI4NET board.