Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
26
Document no. M1125635-05
To remove the module mother board
1. Remove the CPU mother board.
2. Remove the six screws from the front side of the module mother board and one screw
from the bottom of the frame housing.
3. Pull the metal chassis out of the frame housing.
4. Detach the module mother board by removing the
four screws.
Reassemble the device in reverse order.
NOTE: While reassembling, make sure the module connectors are located exactly in the
middle of the openings of the frame housing.