Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
12
Document no. M1125636-05
Defibrillation Sync (pin1)
Digital defibrillator output synchronization signal. Defibrillation Sync indication is generated by
ECG. When active, the signal is in state 1. After 10 ms the signal is reset to state 0. Defibrillation
Sync is not generated before the indication is deactivated. The delay from the R wave peak to
the start of the signal is maximally 35 ms.
Pressure out (pin 7):
P1 from hemodynamic module
The Invasive pressure output signal is 1 V/100 mmHg, originally ranging from 0 to 320 mmHg,
and with a delay of approximately 25 ms. The signal requires an input impedance of 100 k.
Direct ECG (pin 8):
Delay (max.): 15 ms
Gain ECG (out)/ECG (in): 1 V/1 mV
Pacer: 5 V and 2 ms pulse
Table 3 Module connector out (X1-X5), F-CU5(P)
25 pin male D-connector Pin Signal
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Reset_485
-15V
15Vdirty
+15V
Data_485_
Data_485
GND
Reset_485_
CTSB
RTSB
RXDB
TXDB
GND
NC
GND
CTSC
RTSC
RXDC
TXDC
NC
PWM_ECG
RXD_RS232
TXD_RS232
+5V
+5V