Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
22
Document no. M1125636-05
NOTE: Make sure that the cables are properly attached and secured.
• Install the display together with the Command Bar, the Hemodynamic module and E-REC.
Connect the mains power cord and turn the monitor on.
"
23. Fan
Check that the fan is running.
"
24. Starting
Check that the monitor starts up properly, i.e. the alarm LEDs on the Command Bar turn
blank, the start-up sound is heard from the loudspeaker and the normal monitoring
screen appears. No error messages should appear on the screen.
"
25. Module communication
Check that the connected modules are recognized, i.e. the needed parameter information
is shown on the screen and the E-REC records two lines of start-up information.
If some parameter information is missing, check the screen configuration from the
MONITOR SETUP menu.
NOTE: InvBP waveforms are not shown without a patient simulator.
"
26. Real time clock
Check that the clock on the screen shows correct time. Adjust the time, if necessary.
Monitor Setup - Time And Date
NOTE: If the clock shows time 0:00 continuously (at successive start-ups), the
SRAM/TIMEKEEPER chip on the CPU board, or its battery, needs to be replaced.
The FACTORY RESET should be performed after the replacement has been made.
"
27. Loudspeaker sound
Check the loudspeaker by setting the alarm sound:
Alarms Setup - Alarm Volume
Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the
alarm volume changes correspondingly. The alarm sound should be clear and audible
with all the settings.
"
Preset InvBP and ECG measurement settings:
Invasive Pressures - P3 Setup - Label - Art
ECG - ECG1 LEAD -I
Connect a patient simulator to the Hemodynamic multiparameter module.