Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
28
Document no. M1125636-05
To remove the F-CU5(P) frame EMC cover and fan unit
1. Remove the PSM Connection Unit.
2. Remove the six screws (T10) and the hexagon
spacer (size 5.5) to remove the EMC cover.
3. Detach the connector connecting the power board
and module mother board together.
4. Disconnect the fan cable from the power board.
5. Now you can access the following field replaceable
parts:
- Frame unit fan
- EMC cover
- Connection flex board

Table of Contents