Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
34
Document no. M1125636-05
Figure 15 Monitor troubleshooting, page 2
Pick up next module
Clock etc
appears on the
screen?
Plug in a module
Do the
parameter fields
appear on the
screen?
Another
module works in
the same
place?
Faulty module. Go to
Module Troubleshooting
Faulty CPU board. Go to Digital
Section troubleshooting
No
No
Faulty F-CU5(P) Module Frame.No
Yes
Start-up
texts appear on
screen?
Continues from previous page
Switch the monitor
(Stby-switch) ON
Is the
F-CPU fan
running?
Problem in F-CPU power
board; Connect an other board
and schwitch the monitor ON
No
Yes
Does
the CPU board
contain SW?
No
Download appropriate monitor
SW
No
Yes
Yes
Is the
F-CU5(P) fan
running?
Replace the F-CU5(P) power
board; Check connections
No
Check the frame voltages
on Service Page
Yes
Yes
Are
the voltages
OK?
Replace the F-CU5(P) power
board; Check connections
No
Yes
f-cu5_trbl_frameslot1.vsd
Yes