3 Service procedures
3.1 General service information
Field service is limited to replacing faulty PC boards or mechanical parts. The PC boards are
then returned to GE Healthcare for repair.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation and a detailed description of the fault.
WARNING Only trained personnel with appropriate equipment should perform the
tests and repairs outlined in this section. Unauthorized service may void
warranty of the unit.
3.2 Service check
These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (“APPENDIX A:”) which may be used when performing the
procedures.
The symbol
"in the instructions indicates that the check form contains space to record the
results of the particular procedure.
3.2.1 Recommended tools
3.2.2 Visual inspection
If the monitor is switched on, turn the monitor to STBY.
1. Cable
Disconnect the command bar cable from the B-CPU6 board.
Check that the connector pins of the cable are clean, straight and at about the same
height. Check that the cable is intact. Check that the locking screws inside the connector
case are intact.
Leave the cable disconnected.
"
2. Front cover and front panel sticker
Check that the plastic front cover and the front panel sticker are intact.
"
Tool Order No. Notes
Central Unit
Screwdriver