Datex-Ohmeda S/5 Anesthesia and Critical Care Monitor
8
Document no. M1125641-03
2. External parts
Check that:
the front cover and the front panel sticker are intact
the module bus connector is intact
the module box, the latch and the spring pin are intact
the extension module cable and its connector are intact
the screw on the cable connector is tightened properly
"
Reattach the module box and check that the latch is moving properly.
Extension Frame
3. Plastic frame
Check that the plastic frame is intact.
"
4. Mounting plate
Check that the extension frame mount is tightened properly to the plastic frame.
"
5. Fastening screw
Check that the large fastening screw at the back of the mount is intact.
"
6. Pads
Check that all four rubber pads are in place on the bottom.
"
7. Fan filter
If the extension frame contains a fan (Rev. 02->), clean or replace the fan filter.
"
8. Cable connector
Check that the cable connector on the rear panel is clean and intact.
"
9. Module motherboard connectors
Check that the module motherboard connectors are clean and intact. Check also that the
screws that connect the module motherboard to the frame are tightened properly.
"