Datex-Ohmeda S/5 Anesthesia and Critical Care Monitor
8
Document no. M1125641-03
2. External parts
Check that:
− the front cover and the front panel sticker are intact
− the module bus connector is intact
− the module box, the latch and the spring pin are intact
− the extension module cable and its connector are intact
− the screw on the cable connector is tightened properly
"
• Reattach the module box and check that the latch is moving properly.
Extension Frame
3. Plastic frame
Check that the plastic frame is intact.
"
4. Mounting plate
Check that the extension frame mount is tightened properly to the plastic frame.
"
5. Fastening screw
Check that the large fastening screw at the back of the mount is intact.
"
6. Pads
Check that all four rubber pads are in place on the bottom.
"
7. Fan filter
If the extension frame contains a fan (Rev. 02->), clean or replace the fan filter.
"
8. Cable connector
Check that the cable connector on the rear panel is clean and intact.
"
9. Module motherboard connectors
Check that the module motherboard connectors are clean and intact. Check also that the
screws that connect the module motherboard to the frame are tightened properly.
"