System installation
41
Document no. M1137263-04
Scavenging through the ventilator reservoir
Figure 20 Scavenging through ventilator reservoir
1. Connect an exhaust line to the sample gas outlet on the module’s front panel.
2. Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir
tube diameter is at least 2 - 3 times larger than the exhaust line.
Scavenging through the anesthesia gas scavenging system
Anesthesia machines are equipped with an anesthesia gas scavenging system (AGSS), and in
some machines the sample gas outlet can be connected directly to that.
For example, connect the sample gas outlet to the S/5 Avance:
Figure 21 Connecting the gas module to the scavenging connector of S/5
Avance
NOTE: Refer to the anesthesia machine’s documentation to find out where and how the sample
gas can be connected.