Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
54
Document no. M1137263-04
Figure 27 An example of interfacing external devices with Device Interfacing
Solution
(1) Datex-Ohmeda S/5 Anesthesia Monitor (with software L-ANE02(A) or later)
(2) Aestiva/5 anesthesia machine
(3) RGM monitor
(4) Interfacing module
NOTE: You can connect up to ten (10) interfacing modules to one system simultaneously.
Check the maximum number of modules: one meter cable = max. four ten modules, three
meter cable = max. three modules, six meter cable = one module.
WARNING Make sure that the interfacing module is always used in vertical position to
prevent water from entering the module.
WARNING Make sure that you are connecting the interfacing module to the device
specified in the label. Always verify the compatibility of the software
versions before use.
4.2.4 Selecting the external device
1. Turn off the monitor.
2. Turn off the external device.
3. Connect the interfacing module to the monitor’s UPI4NET board or to the connector for
N-DIS or to another interfacing module.
4. Connect the device specific cable to the external device and turn the external device on.
5. Turn the monitor on. The monitor identifies the connected device automatically.