Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
B4(6)
Document no. M1137263-04
(1
For impedance RESP measurement. For invasive pressure measurement in M-series modules except M-PRE(S)TN. The
invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is
guaranteed for all transducers.
(2
For BIS measurement in E-BIS/M-BIS and RESP measurement in M-(P)RE(S)TN and invasive pressure measurement in
M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive pressure measurement compliance level is dependent on
the transducer used, and only the lower level is guaranteed for all transducers.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the S/5™ AM or CCM is used exceeds the applicable
RF compliance level above, the S/5™ AM or CCM should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the S/5™ AM or
CCM.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m or 1 V/m
(1
.