Appendix B: Measuring Floor Flatness
-The floor must meet strict flatness specifications. The information in this appendix is provided as
a tool for accurate measurement of the floor flatness.
Required Tools
Self-leveling fan beam laser tool (self-leveling for at least 3 degrees)
Masking tape
Chalk line
1 m (3') level with minimum 1 mm (1/16") gradations (alternatively, use a tape measure
securely taped to a spirit level)