General System Requirements 1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 1-12
Table 1-2: D670 Additional CT-specific Components and Clearance — Metric
Crated Uncrated
Component
name
Crate size
(Dollies not attached)
(cm) (Height × Width ×
Depth/Length)
Weight (kg)
Minimal dimensions (cm)
*
Weight (kg)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90
° turns
†
Height
CT PDU 123×90×70
D670/OPT 460
D670/BSE 440
100 100 100 100 any
D670/OPT 370
D670/BSE 350
CT UPS
82×31×81 any any any any any 160
CT console (NIO)
65×47×83 115 any any any any any 88
CT accessories
115×100×150 50 100 110 150 180 any 50
CT sub-system
covers
190×100×170 537 80 80 100 128 199 30
MDP (Main
Disconnect Panel)
‡
**
100×70×50 60 any any any any any 45
Optional Items
Contrast Injector May vary but not
more than
80×80×80
May vary but not
more than 15 any any any any any <13
AW W/S
* The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor leading to
the room. When planning or measuring the width of the scan room door, use the graphs provided in Figure 1-2: D670 Relative Required Width for
Corridor and Scan Room Door, p.1-17 in order to verify that the measurements comply with the requirements.
† The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See Figure 1-3: D670 Required Corridor Width for 90° Turns,p.1-18 for the required width when the angle is 90°.
‡ A1 or other; if ordered from GE, delivered in advance as part of the site preparation and must be installed prior to system installation
** In some cases, an alternative MDP is supplied, in order to comply with local electricity standards, see Power Input Distribution Box (PIDB) with MDP
(A1), p.5-14