General System Requirements 1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 1-13
Table 1-3: D670 Components and Clearance — Imperial
Crated Uncrated
Component name
Crate size (")
(Dollies not attached)
(Height × Width ×
Depth/Length)
Weight
(lb)
Minimal dimensions (")
*
Weight
(lb)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90
° turns
†
Height
Pre-installation kit
‡
29.5×5.7×68.9 33 any any any any any 33
NM gantry
86.6×59×66.1 4917 55.1 55.1 88.6 98.4 78.75
4531
(with dollies)
Table 55×35.4×118.1 1625 39.4 39.4 111.4 98.4 any 1245.6
NM acquisition station
31.5×23.62×23.62 66 any any any any any 44
Peripherals and accessories 45.3×39.4×59 110 any any any any any 110
Collimators on carts
67×35.4×45.3
816
(heaviest
coll. set)
22 22 39.4 45.3 59
727.5
(heaviest coll.
set)
CT gantry 91x54x78.75 4740 45.3 45.3 110.6 98.4 78.75
4432
(with dollies)
NM/CT covers 81.5×41.3×77.6 198 any any any any any
55
Optional Items
NM UPS
May vary but not
more than
23.6x15.7x31.5
May vary
but not
more
than
330.7
any any any any any
May vary
but not
more than
286.6