EasyManua.ls Logo

GE Discovery NM/CT 670 - Floor Vibration; Floor Conductivity Recommendations; Additional Floor Requirements

GE Discovery NM/CT 670
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Equipment Description and General Construction Requirements 2.3 Room Structural Requirements
2.3.1 Floor Requirements
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 2-41
2.3.1.5 Floor Vibration
Floor vibration requirements are included in the general vibration requirements
(see Vibration Specifications, p.2-44).
2.3.1.6 Floor Conductivity Recommendations
The purpose of this section is to measure the electrical conductivity of the floor surface to the
“GND” (Ground).
The surface of the conductive floor shall provide a patch of electrical conductivity between all
persons and equipment making contact with the floor.
Using a DVM, measure the impedance between the upper surface of the floor – where the
NM gantry is planned to be positioned, and the system power supply GND terminal in the
room. The readout needs to be <35 M Ohm.
Repeat the measurement in the area where the patient table will be positioned. The readout
needs to be < 35 M Ohm.
Repeat the measurement in the area where CT gantry will be positioned. The readout needs
to be < 35 M Ohm.
2.3.1.7 Additional Floor Requirements
The floor finish must take into consideration magnetic field and EMI considerations
(see Special Construction Requirements, p.3-1 andEMI Considerations,p.3-7).

Table of Contents

Related product manuals