K415 Issue No. 1 8
Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the
units required, press TASK and select any task, other than BASIC, press SETUP and
proceed as detailed on page 37. To return to BASIC mode, press TASK and select BASIC.
In BASIC mode, the instrument is configured to carry out basic Pressure to Voltage (P to
V) or Pressure to Current (P to I) tests, a typical test procedure follows.
Voltage and Current Measurement
Connect the electrical input sockets as follows for voltage and current measurements.
Use the test leads provided and DO NOT push bare wires into the sockets.
Note: Maximum applied voltage = 50V dc,. Maximum input current = 55mA dc
Note: Maximum applied voltage for external loop supply = 30V dc
SELECTIONNER
UNITES
inHg
bar
hPa
Pa
bar
V
COURANT
UNITES
FCT.: STANDARD
CHOIX UNITES
DE PRESSION
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
+
-
V
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
+
-
Getting Started