K415 Issue No. 1 12
Pressure Transmitter (P-I) Task
Select the P-I task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below.
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure to current 24V device
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Voltage Output Pressure Transmitter (P-V) Task
Select the P-V task from the task menu and connect the UUT to the calibrator as shown
below. Voltage output transducers with a 10 V supply and outputs of ±10 V can be
calibrated using the 10V output sockets (supply) and the transducer output connected to
a voltmeter (refer to Method on page 13 for a test procedure).
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Excitation 10V
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
P
I
+ (24V)
-
bar
mA
F2F1
A
B
FCT. : P-I 24V OFF
MODE ENREG. RAPIDE
DPI 615
C
F2F1
P
V
-+
+
-
+
-
bar
V
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
A
B
C
Taking Measurements