K415 Issue No. 1 16
Pressure to Display (P - Display) Task
P-Display is a special application of Datalog. To use this mode, select Datalog from the
Store Mode menu as detailed on page 37. Connect the UUT to the instrument as shown
below and, if required, turn on and set-up Cal Mode (see page 11).
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Dial gauge under test
Press
TASK
and select
P-DISPLAY
. If required, use
OUTPUT
key to change pressure units.
Set-up a data log file as detailed on page 31.
Note: TRIGGER field, automatically set to KEYPRESS, cannot be changed.
Apply a series of test pressures to the device under test. Enter displayed reading at
each pressure and log each point:
After logging final test point, terminate as
follows:
bar
F2F1
AFFICHAGE
CHANGER
VALEUR
A
B
C
FCT. : P-AFFICHAGE 24V OFF
FICHIER : AUCUN
DPI 615
ENTRER VALEUR
VALEUR
CHANGER
VALEUR
VALEUR
VALEUR
STORE
+
bar bar
bar
FCT.: P-VALEUR
FICHIER: A1
FCT.: P-VALEUR
FICHIER: A1
FCT.: P-VALEUR
FICHIER: A1
24V OFF 24V OFF 24V OFF
CHANGER
VALEUR
ENREGISTRER
MAINTENANT
ARRETER
ENREGISTREM.
AFFICHAGE
bar
ENREGISTRER
MAINTENANT
ARRETER
ENREGISTREM.
DERNIER POINT
FCT.: P-AFFICHAGE 24V OFF
:A1
FICHIER
STORE
Taking Measurements

Table of Contents

Related product manuals