K415 Issue No. 1 18
Transmitter Simulation (TX SIM) Task
Provides a current output proportional to the calibrator’s measured output pressure
(indicated pressure on indicator only version). Select task TX SIM. Press EXIT to skip set-
up screen if parameters are correct.
On completion of TX SIM set-up, the display is configured as follows:
Set-up the output loop as detailed on page 25 and, if necessary, turn on the internal
electrical supply.
To subsequently change any of the TX SIM scaling parameters, press CHANGE VALUE
(F1) to obtain the TX Simulation display.
To change the pressure units, press INPUT and select the required scale units. If the
required scale units are not listed, press SETUP, select SETTINGS and proceed as detailed
on page 38.
REGLER VALEUR
TRANSMETTEUR
VALEUR MINIMUM
kPa0.000
mA4.000
VALEUR MAXIMUM
kPa
20.000
mA
20.000
VALEUR MINIMUM
kPa0.000
mA4.000
VALEUR MAXIMUM
kPa
20.000
mA
20.000
VALEUR MINIMUM
kPa0.000
mA4.000
VALEUR MAXIMUM
kPa
mA
20.000
VALEUR MINIMUM
kPa0.000
mA4.000
VALEUR MAXIMUM
kPa
20.000
mA
20.000
20.000
TRANSMETTEUR
TRANSMETTEUR
TRANSMETTEUR
REGLER VALEUR
REGLER VALEUR
REGLER VALEUR
CHANGER
VALEUR
CHANGER
VALEUR
CHANGER
VALEUR
CHANGER
VALEUR
PRESSURE INT
bar
mA
CURRENT OK
TASK : TX SIM
SNAPSHOT MODE
CHANGE
VALUE
24 V OFF
Taking Measurements