K415 Issue No. 1 26
mA Step
To select one of the electrical output programs, press the OUTPUT key and proceed as
follows:
On selection of (e.g.) Linear, the output display window changes to show the selected
program of output currents:
• For current sourcing applications, turn on the 24V supply as shown on page 28. For
current sinking applications, connect an external supply as shown on page 25 and
leave the 24V supply switched OFF.
•Press RUN (F1) to run program. A flashing status display CHECK LOOP indicates a
fault in the external loop i.e. supply fault or open circuit.
Note: The dwell time at each step is approximately 10 seconds.
•Press STOP (F1) when running to stop at any point. Press RUN (F1) to resume.