31 K415 Issue No. 1
Data Log Operations
Data Log is a special application of store mode that enables the calibrator to either
automatically log displays at preset time intervals or to manually log a display on
operation of the STORE key. Logged data is written to a user specified file.
To set-up a Data Log file, proceed as follows:
• Select a task, other than BASIC. If using ADVANCED, set-up required output
parameters.
•Use SETUP to select Data Log from the Store Mode Menu (see page 37).
Auto Log (Timer)
Press STORE and set up the Data Log file parameters as shown below. Use CHANGE
VALUE (F1) followed by cursor keys to set field values. For Auto Log, set TRIGGER field to
PERIODIC.
Manual Logging
Enter the file details as shown above and select KEYPRESS for TRIGGER field. Screen
reverts to displayed parameters showing set-up file as shown below:
Use a combination of STORE and LOG NOW (F1) to log events as follows:
ENTER VALUE
STORE
LOGFILE2
PRESSURE INT
bar
VOLTAGE
V
CHANGE
VALUE
DATA LOGGING
FILENAME :
DATE :
TIME :
TRIGGER:
PERIOD :
POINTS :
21/02/99
12:29
00:05
10
PERIODIC
CHANGE
VALUE
DATA LOGGING
FILENAME :
DATE :
TIME :
TRIGGER :
PERIOD :
POINTS :
21/02/99
12:29
00:05
10
PERIODIC
START
START
LOGGING
LOGGING
(ANY KEY)
CLEAR
TASK : CO2INLET
POINT 10 OF 10
24 V OFF
CHANGE
VALUE
DATA LOGGING
FILENAME :
DATE :
TIME :
TRIGGER :
PERIOD :
POINTS :
21/02/99
12:29
KEYPRESS
LOGFILE3
CHANGE
VALUE
DATA LOGGING
FILENAME :
DATE :
TIME :
TRIGGER :
PERIOD :
POINTS :
21/02/99
12:29
KEYPRESS
LOGFILE3
PRESSURE INT bar
VOLTAGE V
LOG FILE : LOGFILE3
TASK: ADVANCED
START
LOGGING
START
LOGGING
24 V OFF
PRESSURE INT bar
ENTER VALUE
PRESSURE INT bar
VOLTAGE V
LOG FILE : LOGFILE3LOG FILE : LOGFILE3
TASK: ADVANCED
TASK: ADVANCED
STORE
VOLTAGE
V
LOG
NOW
STOP
LOGGING
24 V OFF
24 V OFF
Memory Operations Data Log