K415 Issue No. 1 36
Recalling Data Files (DPI 615 instruments only)
Data or results files generated by running procedures are stored in the instrument’s Data
Log directory. To recall a data file to the display, ensure that DATA LOG is selected from
the SETUP menu and proceed as follows:
Use the cursor keys to select either the AS FOUND option or the AS LEFT option for
display. AS FOUND is the result of the first run of a procedure and AS LEFT is the result of
the last time the procedure was run.
Procedure data files can be displayed either as text (stored screens) or in graphical form.
To display as text, select the TEXT option (F1) from the directory and proceed as follows
from the File Summary menu. Select AUTO STEP (F1) to automatically review each screen
at 1 second intervals or use the left or right cursor keys to manually review.
To transmit the selected logged data via the RS232 interface, connect the instrument to
a free port on an external PC, ensure that the instrument’s RS232 parameters match
those of the PC.
To display logged data in graphical form, on screen, select GRAPH (F2) from the directory
and proceed as follows:
bar
V
TASK:P-V
DATA LOG
AUTO
STEP
24V OFF
. %span FAIL
DATA POINT
5OF10
MANUAL REVIEW
TRANSMIT
FILE
REVIEW
LOGGED DATA
USER ID:
SERIAL No.
RESULTS
CLIFF
99127007
FAIL
FEEDPRESS11
MAX PRESS:
POINTS:
MAX. ERROR:
20
5 UP/DOWN
0.1% SPAN
DIRECTORY
TEXT
GRAPH
AS FOUND
AS LEFT
FEEDPRESS11
V
time
0
10.007
15:58:55
16:18:23
Memory Operations Data Log