K415 Issue No. 1 46
Internal Pressure Range
Use the following procedure for calibrating the internal pressure range.
Note: If calibrating the hydraulic calibrator version, the calibrator must first be primed as
detailed on pages 59 to 66.
(1) Connect the outlet port of the instrument to a pressure standard.
(2) Allow the instruments temperature to stabilize for a minimum of 1 hour.
(3) Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu and then PRESSURE INT.
(4) Apply zero pressure and store the zero point:
(5) Close the vent valve, apply full-scale pressure and store the full-scale point:
(6) Press the ENTER key to accept the calibration. Press the EXIT key four times to
quit CALIBRATION and SETUP modes.
(7) Check calibration by applying test pressures in Table 1, page 44.
VOLTAGE IN
Calibration Terminee
App ENTER pour continuer
Memoriser point PE ?
OUI NON
m = 2.784670e-04
c = -1.459980e+02
Applique Mesure
0.0000 524293
20.000 704637
CALIBRE PRESSION INT
GAMME 20.000 bar g
CALIBRE PRESSION INT
GAMME 20.000 bar g
RESULTATS
Calibration