53 K415 Issue No. 1
Current Output Range (24 mA)
Use the following procedure for calibrating the current output range.
(1) Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION MENU as shown on page 45.
(2) Select CURRENT OUT range for calibration:
(3) Connect a digital milliammeter to the instrument as shown below.
Note: On standard instruments, during a current out calibration routine, the 24V
dc output is automatically turned on.
mA
Calibration