K415 Issue No. 1 56
Ambient Temperature Channel
Use the following procedure for calibrating the ambient temperature measurement
channel.
(1) Switch the instrument on, enter the CALIBRATION mode and select
TEMPERATURE:
(2) Allow the instruments temperature to stabilize in the calibration environment for
at least one hour.
(3) Read the environmental temperature on a calibrated digital thermometer and
enter the recorded value as shown below. Example shown for a measured
ambient temperature of 21.5° Celsius.
Note: Only one temperature point is required.
(4) Press the ENTER key to accept the calibration. Press the EXIT key once to return
to the calibration menu or four times to quit the CALIBRATION and SETUP modes.
CHANGER
VALEUR
DEMARRER
Entrer val appliquee:
CALIBRE TEMP AIR
-10 à 40°C
CHANGER
VALEUR
Entrer val appliquee:
+
CHANGER VALEUR
Etalonnage Accepte ?
OUI NON
CALIBRE TEMP AIR
-10 à 40°C
CALIBRE TEMP AIR
-10 à 40°C
m = 2.49000e-04
c = -1.359980e+02
Applique Mesure
21.5 623298
CALIBRE TEMP AIR
-10 à 40°C
RESULTATS
Calibration Terminee
App ENTER pour continuer
Calibration