K415 Issue No. 1 76
Apply a nominal zero pressure to the instrument and store the zero point:
Apply the positive, half full-scale pressure to the instrument and store the
positive, half full-scale point:
Apply positive full-scale pressure to the instrument and store the full-scale point:
After application of the final calibration pressure, the calibration results are
displayed as shown below. To view the second page of calibration data press
MORE (F1). Press ENTER to complete the pressure calibration procedure.
Entrer val appliquee:
Appliquer zero nominale
App sur ENTER quand la
la valeur est stable
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
NON
OUI
CHANGER
VALEUR
REGLER VALEUR
Memoriser point ?zero
Entrer val appliquee:
Appliquer 50% valeur +ve
App sur ENTER quand la
la valeur est stable
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
NON
OUI
CHANGER
VALEUR
Valider 50% point +ve?
REGLER VALEUR
Entrer val appliquee:
Appliquer PE nominale
App sur ENTER quand la
la valeur est stable
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
NON
OUI
CHANGER
VALEUR
Memoriser point P.E.?
REGLER VALEUR
RESULTATS
RESULTATS
a= 1.632719e-13
b= -7.546076e-06
c= 3.915159e+00
Applique Mesure
0.0000 523166
1.2500 354205
2.5000 182699
Calibration Terminee
App ENTER pour continuer
a= 1.47115e-14
b= -7.546076e-06
c= 3.808651e+00
-2.5000 865995
-1.2500 694638
0.0000 523166
Applique Mesure
Calibration Terminee
App ENTER pour continuer
POURSUIVRE POURSUIVRE
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L