User Manual Druck DPI620 Genii
4. Read the CH2 value and enter this in the meter value text
box using the keypad.
Figure 6-9 Enter Calibration Point
5. Select SET.
6. Repeat steps 3 and 4 with 20 mA selected. This will
calibrate the transmitter’s output current.
06 June 2013 K0541 issue 1 Page 106 of 175