User Manual Druck DPI620 Genii
Electrical warnings 1.6
• To prevent electrical shocks or damage to the instrument,
do not connect more than 30V CAT I between the
terminals, or between the terminals and the ground
(earth).
• External circuits should have appropriate insulation to
the mains.
• To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages
that are more than 20 Vrms. Do not connect more
than 300V CAT II between the IO620-AC leads, or
between the leads and the ground (earth). Attach it to
the specified connections only.
• This instrument uses a Lithium-Polymer (Li-Polymer)
battery pack. To prevent an explosion or fire, do not
short circuit, do not disassemble, and keep it safe
from damage.
• To prevent an explosion or fire, use only the GE
specified battery (Part: 191-356), power supply (Part:
191-339) and battery charger (Part: IO620-CHARGER).
• To prevent battery leakage or heat generation, only
use the battery charger and power supply in the
temperature range 0°C to 40°C (32°C to 104°F).
• The power supply input range is 100 – 240Vac, 50 to 60Hz,
250mA, installation category CAT I.
• Position the power supply so not to obstruct the supply
disconnecting device.
06 June 2013 K0541 issue 1 Page 12 of 175