User Manual Druck DPI620 Genii
To do a calibration on a measure or source function, use the
advanced menu option.
1. Select:
DASHBOARD >> ADVANCED
2. Enter the calibration PIN: 4321
3. Select the button.
4. Select:
PERFORM CALIBRATION
Then select a function and start the calibration:
1. Select channel
2. Select function
3. Select range (if applicable)
4. Follow on-screen instructions
When the calibration is complete, set the next calibration
date.
Procedures (CH1/CH2): Current (measure) 8.2
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
06 June 2013 K0541 issue 1 Page 142 of 175