User Manual Druck DPI620 Genii
Procedures (CH1): DC mV/Volts (source) 8.5
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration (Zero and +FS) for the applicable range:
mV (source) ranges Volts (source) ranges
2000 mV 20 V
4. Check the calibration is correct:
• Select the applicable Millivolts or Voltage (measure)
function (ref Section 0). Apply the input values that are
applicable to the calibration:
mV: 0, 100, 200, 1000, 2000
Volts (V): 0, 5, 10, 15, 20.
• Check the error is within limits (Ref: Table 8-6 or Table
8-7)
06 June 2013 K0541 issue 1 Page 148 of 175