User Manual Druck DPI620 Genii
Frequency calibration (source function)
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Set-up the equipment as follows:
Frequency meter: Gate time = one second
Druck DPI 620 Genii: Waveform = Square;
Amplitude = 10 V
Frequency = 990 Hz
4. Use the calibration menu (Ref: Section 8.1) to do a one-
point calibration.
5. Check the calibration is correct.
06 June 2013 K0541 issue 1 Page 151 of 175