User Manual Druck DPI620 Genii
Frequency calibration check
Frequency (measure) calibration check.
Signal generator: Output = 10 V
Unipolar
Square wave
Druck DPI 620 Genii: Input trigger level = 5 V
Units: Hz or kHz as specified in
Table 8-8 or Table 8-9
Frequency (source) calibration check.
Frequency meter: Gate time = one second
Druck DPI 620 Genii: Units: Hz or kHz as specified
in Table 8-8 or Table 8-9
Select the applicable Millivolts or Voltage (measure)
function (ref Section 2.10). Apply the input values:
Hz: 0, 990
kHz: 10, 50
Check the error is within limits (Ref: Table 8-8 or Table
8-9).
06 June 2013 K0541 issue 1 Page 152 of 175