User Manual Druck DPI620 Genii
5. Check the calibration is correct.
Set-up the equipment as follows:
Source frequency = 0 (For DC output);
Waveform = Square
Apply the amplitude values that are applicable to the
calibration (Ref: Table 8-10).
Check the error is within limits (Ref: Table 8-10).
Table 8-10 Amplitude (source) error limits
Amplitud
e
Volts (V)
Calibrator
uncertainty
(V)
Permitted
DPI 620
error (V)
0.2
0.01
0.1
5.0
0.01
0.1
10.0
0.01
0.1
20.0
0.01
0.1
06 June 2013 K0541 issue 1 Page 155 of 175

Table of Contents