User Manual Druck DPI620 Genii
CJ (TC Method)
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Set the reference unit temperature: 0°C.
3. Let the equipment get to a stable temperature
(minimum: 1 hour since the last power on).
4. Use the calibration menu (Ref: Section 8.1) to do a one-
point calibration for the CJ (TC method) function.
5. Calculate the expected reading, using the known error in
the thermocouple and reference unit.
6. Check the calibration is correct.
Select the TC measure function (Ref: Section 2.12).
Check the equipment gives a TC temperature that is
the same as the temperature on the reference unit
±0.1°C (0.2°F), after correction for the known
thermocouple and reference unit error.
06 June 2013 K0541 issue 1 Page 163 of 175