User Manual Druck DPI620 Genii
CJ (TC Method)
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Set the reference unit temperature: 0°C.
3. Let the equipment get to a stable temperature
(minimum: 1 hour since the last power on).
4. Use the calibration menu (Ref: Section 8.1) to do a one-
point calibration for the CJ (TC method) function.
5. Calculate the expected reading, using the known error in
the thermocouple and reference unit.
6. Check the calibration is correct.
• Select the TC measure function (Ref: Section 2.12).
• Check the equipment gives a TC temperature that is
the same as the temperature on the reference unit
±0.1°C (0.2°F), after correction for the known
thermocouple and reference unit error.
06 June 2013 K0541 issue 1 Page 163 of 175