User Manual Druck DPI620 Genii
4. Check the calibration is correct:
• Select the applicable AC mV or AC Volts (measure)
function (ref Section 0).
5. Apply the input values that are applicable to the
Calibration.
• AC mV: 10, 500, 1000, 2000
• AC Volts: 5, 10, 20
6. Check the error is within limits (Ref: Table 8-15 or Table
8-16)
Table 8-15 AC mV (measure) error limits
AC mV
uncertainty
(mV)
DPI 620 error
(mV)
Table 8-16 AC Volts (measure) error limits
06 June 2013 K0541 issue 1 Page 166 of 175