Procedures: Pressure Indicator Modules (PM 620) 8.14
Do the procedure as follows:
1. Assemble the pressure indicator with the necessary
PM620 modules.
2. Connect the instrument to the pressure standard.
3. Let the equipment get to a stable temperature
(minimum: 60 minutes since the last power on).
4. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration.
Zero and +FS for absolute sensors.
Note: If software version requires a three-point
calibration for an absolute sensor, use points at 0, 50%
and +FS or three-point calibration:
-FS, Zero and +FS for gauge sensors.
(Ref: Table 8-17).
06 June 2013 K0541 issue 1 Page 167 of 175