5. Check the calibration is correct:
Select the applicable pressure function.
Apply the following pressure values (absolute sensors).
0, 20, 40, 60, 80, 100 (%FS).
Go back to 0 in the same steps.
Apply the following pressure values (gauge sensors.
0, 20, 40, 60, 80, 100 (%FS).
Go back to 0 in the same steps.
Check the error is within limits (Ref: Table Gauge
Ranges and Absolute Ranges).
Gage ranges or absolute ranges tables (Ref:
datasheet.
Use the values in the total uncertainty column.
The specified values include an allowance for
temperature changes, reading stability for one year,
and the uncertainty of the standard used for
calibration.
06 June 2013 K0541 issue 1 Page 169 of 175