User Manual Druck DPI620 Genii
2.14.2 IDOS Function Procedures
Set the calibrator functions for use on the display.
This includes:
• IDOS function (to measure the pressure)
• If necessary, change the Units for the function
• If necessary, set a Utility for the function:
i Max/Min/Avg
ii Switch Test
iii Leak Test . The procedure is the same for an
IDOS module or for a MC 620/PM 620
assembly.
• If necessary, change the Settings for the IDOS function:
• Process (Tare, Alarm, Filter, Flow, Scaling)
i Leak Test (Only when the Utility is set). The
procedure is the same for an IDOS module or for
a MC 620/PM 620 assembly.
ii Zero. The procedure is the same for an IDOS
module or for a PM 620 module. Zero the
gauge sensor before use.
• Continue with the pressure operation.
06 June 2013 K0541 issue 1 Page 62 of 175