User Manual Druck DPI620 Genii
Step Procedure (refer to Figure 3-5)
1
Remove the adaptor from the pressure port.
2
Use an applicable seal for the pressure connection:
a. NPT type: Use an applicable sealant on the
thread.
b. BSP (parallel) type: Use the applicable
bonded seal at the bottom.
c. BSP (parallel) type, 100 bar (1500 psi) or
less: a bonded seal at the top is permitted.
3
Attach the adaptor to the external equipment. If
necessary use an alternative adaptor.
4
Tighten to the applicable torque.
5
Attach the adaptor to the MC 620 carrier and hand-
tighten.
When the pressure indicator assembly is complete use the
menus to set-up the necessary operations. (refer to Section
3.3).
06 June 2013 K0541 issue 1 Page 69 of 175