User Manual Druck DPI620 Genii
6 HART® Operations
The Druck DPI 620 Genii can communicate with devices that
use the HART® protocol as follows:
The Universal and Common Practice commands
specified in HAR revision 5 to 7.
Devices that support Device Descriptions (DD).
This section includes procedures to use the HART functions
available in the calibrator.
HART® Menu Operations 6.1
The HAR application is started from the Dashboard. The
HART® uses a digital signal on top of a standard 4 - 20 mA
current loop to get data to and from a HART® enabled field
device. Typical operations include the following:
Read the primary variable and the analogue output.
Read the device serial number, type and supplier.
Get calibration data (upper and lower range values,
sensor limits, calibration date).
Do status and fault finding checks.
Change the device configuration (range, units,
damping).