User Manual Druck DPI612
Page 104 of 105 [EN] English – 109M4017 Revision -
General 8.4
connect / write
to USB Memory
Stick
capacity of 1GB can take a
few seconds to fully connect
Try with smaller capacity USB Drive
Unit will not
connect to
calibration
management
software
Communications mode set
to storage
From the DASHBOARD select
SETTINGS then DEVICES; the USB
client port should be set to
Communications.
Unit will not
Switch On
Faulty / Flat battery Try Using External Power Supply
Remove Battery and Replace
Unit
Press and hold power switch
Remove Battery / Power and replace